Main Features
1.Microcomputer control system
“SMART IIII” microcomputer control system is adopted to realized not only intelligent controls of the sterilization process but also the classified management of instruments (user, administrator and engineer three levels included).As a result, we can guarantee satisfactory use and management of autoclaves.
2.Multilayered pressure protection system
Our patented pressure interlocking device can automatically sense the pressure in the sterilization chamber and tightly locked, which effectively prevent personal injuries caused by spraying steam arising from misoperation.The system can monitor the pressure changes in the sterilization chamber on a real-time basis. Upon detection of any pressure anomaly, the dual overpressure protection system consisting of a safety valve and a pressure protective device will be actuated immediately to relieve pressure or disconnect power supply.
3.Unique anti-scald design
The cover and bench of the sterilization chamber are made of special plastic to give a nice look and avoid causing corrosions and scald. The unique steam discharge and collection system can help to prevent scald.
4.Multiple sterilization cycles and memory system
5 sterilization cycles are preset for sterilization, drying, warming of solid and liquid articles and for melting and warming of agar separately.
5.Temperature and pressure
Sterilization temperature:105 to 135°C,designed pressure:0.3Mpa
6.Big LCD Screen can clearly show all the sterilization data and process
7.Over temperature protection system
The system can monitor the temperature changes in the sterilization chamber on a real-time basis and disconnect power supply immediately upon detection of over temperature or abnormal temperature variation.
8.Automatic troubleshooting system
Troubleshooting is conducted automatically upon startup. Acoustic alarms will be given and Error Number will be displayed on the screen if any trouble occurs during equipment operation.
9.Safety and Function
Self-induction pressure interlocking device, lid closing checking, over temperature protection, temperature monitor, dry scorch protection system, over pressure protection, safety valve, over current and short circuit protection, leakage protection device, anti-scald safety protection, automatic troubleshooting system.
Technical Specifications
Model | AE-STR001 (60L DA) | AE-STR002 (80L DA) | AE-STR003 (110L DA) |
AE-STR001 (60L DF) | AE-STR002 (80L DF) | AE-STR003 (110L DF) | |
AE-STR001 (60L DR) | AE-STR002 (80L DR) | AE-STR003 (110L DR) | |
Model difference | DA:Standard, DF:Automatic water feeding cooling fan, DR: DF drying | ||
Capacity(L) | 60 | 85 | 110 |
Dimension(L*W*H,mm) | 660x644x980(DA)
660x831x980(DF/DR) |
660x644x980(DA)
660x831x980(DF/DR) |
660x644x1180(DA)
660x831x1180(DF/DR) |
Chamber
dimension(Dia*H,mm) |
φ400×505 | φ400×700 | φ400×895 |
Rate power supply(w) | 2900 | 4600 | 4600 |
Stainless steel
basket(Dia*H,mm) |
(φ360×200)x2 | (φ360×280)x2 | (φ360×250)x3 |
Power requirement | 220V,16A,50/60HZ | 220V,32A,50/60HZ | 220V,32A,50/60HZ |
Temperature accuracy | 0.1°C | ||
Working environment | 5-45°C, relatively humidity 10%-85% | ||
Chamber material | SUS 304 | ||
Sterilization temperature | 105-138°C | ||
Sterilization time range | 1-6000 minutes | ||
Melting temperature | 60-100°C | ||
Melting time range | 1-6000 minutes | ||
Warming temperature | 45-60°C | ||
Warming time range | 1-9999 minutes | ||
Drying(only for DR model) | 1-300 minutes, drying temperature 80-140°C | ||
Cooling lock OPEN temperature | 40-80°C for liquid cycle, 40-99°C for other cycles | ||
Sterilization cycle | 1.Solid cycle 2.Liquid cycle 3.Liquid with warming cycle 4.Agar cycle 5.Solid
with drying(for DR) 6. Medical waste cycle 7. Customized cycle |
||
Sterilization cycle number | DA can save up to 20 cycles, DF/DR can save up to 60 cycles | ||
Exhaust mode | Automatic internal discharging ,with a steam collecting bottle inside.
Exhaust speed can be selected by 6 levels |
||
Lid opening way | Up and down | ||
Steam collector and water tank | DA: embedded 2 steam collector, DF: embedded 2 steam collector and 28L
water tank DR: embedded steam collector and 28L water tank |
||
Preset range of timer | Can be preset and auto start up to 6 days in advance | ||
Display range of pressure gauge | -0.1-0.5Mpa | ||
Pressure | Max working pressure:0.27Mpa,design pressure:0.35Mpa | ||
Controller | “Inspiration” fast speed microcomputer controller | ||
Safety device | “eight column evenly distributed” interlocking device, electric double inner locks, lid closing checking, dry scorch protection, overpressure protection, safety valve, water level sensor, over temperature protection, temperature monitor, over current and short circuit protection, leakage protection, cooling lock, anti-scald chamber cover and bench, automatic
troubleshooting |
Standard Accessories
Items | Name | Qty |
1 | Main Machine | 1set |
2 | Operating Manual | 1pc |
3 | Stainless Steel Basket | 2 or 3pcs |
4 | Connect Pipes | 1 packet |
5 | Waste Liquid Bottle | 1pc |
6 | Water Level Plate | 1pc |