Lab companions – Autoclaves (Vertical Type)

Performance

  • Microprocessor controller and temperature calibration function allows for high precision temperature control.
    – Sterilization temperature at 110 to 123°C
    – Melting* temperature at 60 to 100°C
    * Liquefaction of coagulated agar media.
  • Automatic sterilization system for unattended operation.
    – Heating > Check safe > Sterilization > Sterilization safe > Exhausting > End
  • Convenient operation modes for various uses.
    – Standard modes and program modes are listed in detail as below tables.
  • Temperature calibration function.
Standard modes Temp.(°C) Time(min.)
Decontamination 121 35
Solids sterilization 121 20
Liquids sterilization 121 15
Melting 100 30
Program modes Temp.(°C) Time(min.)
Sterilization 1 110 to 123 1 to 999
Sterilization 2 110 to 123 1 to 999
Liquid sterilization 110 to 123 1 to 999
Melting 60 to 100 1 to 999

Convenience

  • Conveniently monitored by an independent control display, sterilization process display, and pressure gauge.
    – VFD for visualization of temp, pressure(optional), timer, warning alert checks.
    – LED lamps for sterilization process check.
    – Easy pressure check with the standard gauge attached to the front.
  • Vertical type autoclaves which allow for top loading are suitable for high volume sterilization.
    – Baskets are stackable two high or even more levels, which secure inside the chamber space requirements.
  • Low-maintenance
    – The chamber is constructed of stainless steel which is superiorly resistant to corrosion.
    – Easy to replace silicone gasket.
  • Convenient documentation with the optional data recorder.
    – Dot type recorder(6 point) allows for precise documentation.

Safety

  • Mechanical safety devices
    – Over current protection.
    – Over temperature protection by automatic power cutoff.
    – Over pressure protection by automatic and manual safety valve.
  • Safety warning systems
    – Over temperature warning.
    – Low temperature warning.
    – Low water level warning.
    – Low heat warning if the temperature does not reach the sterilization or melting temperature.
    – Sterilization fail warning if the operation is unexpectedly stopped during sterilization process.
  • Slide opening door with plastic cover of low thermal conductivity prevents burns from the contact.