This single workstation system is equipped with an integrated gas purification system (one column), PLC controller (with a touchscreen HMI), electronics enclosure, large and small antechambers, replaceable sight glass, and vacuum pump, Attainable purity of this system is H2O < 1ppm, O2 < 1ppm.
This system includes a large capacity purification system capable of adsorbing 60l of O2 and absorbing 2000g of moisture extending your time between regeneration cycles.
Features
- Closed Loop Gas Circulation
Inert gas in a closed loop. The gas is circulated by the blower and purifier, H2O, O2 can be removed continuously. - Auto Purging
The replacement of the atmosphere inside the glove box can be achieved automatically by the purging valves. - Automatic Regeneration
H2O and O2 removal material can be regenerated. The regeneration process can be program controlled. - Automatic Pressure Control
The pressure in the glove box is controlled automatically by the Programmable Logic Controller (PLC). Working pressure can be set either between 15mbar and -15mbar or between 12 mbar and -12 mbar. If the pressure goes over the set range, the system will be protected automatically. - Eco Mode
The vacuum pump will be activated automatically when necessary, and will turn off after a period of idle time. Blower frequency will be switched to 25Hz when moisture and oxygen level reaches to less than 1 ppm
Specifications
- Main Chamber Dimensions
Length: 1220mm (48.0”)
Depth: 750mm (29.5”)
Height: 900m (35.4”)
Material: Stainless steel type 304, thickness 3mm (1/8”)
Sight Windows: Inclined, Safety Glass, thickness 8mm (1/3”)
Gloves Ports: POM (Polyoxymethylene), 220mm (8.7”) Diameter
Gloves: 8B1532
Dust Filter: HEPA 0.3μ, one gas inlet, one gas outlet
Shelving: Stainless Steel, two fixed shelves (adjustable shelves available)
Lighting: External LED Lights, front mounted above sight glass.
Feedthroughs: One power feedthrough, 3 blind-flanges - Large Antechamber
Diameter: 360mm (14.2”) or 400mm (15.7”), Length 600mm (23.6”)
With Sliding tray
Spindle lock door, Vertical operation with lifting mechanism
Automatic operation via solenoid valve - Small Antechamber
Diameter 150 mm (5.9”) or 100mm (3.9”), Length 300mm(11.8”)
Hinge door
Manual operation via hand valve - Purifier
Removal of H2O and O2 to Less than 1 ppm moisture and oxygen
Absorber Unit: 5kg Copper catalyst,5kg Molecular sieve
Capacity to remove oxygen, 60L , moisture 2000g.
Encapsulated Blower: 90m3/h
Regeneration: PLC control all process including heating, introducing mix gas (working gas mixed with Hydrogen 5-10%), etc. - System Control PLC and HMI: Siemens S7
- Vacuum Pump 8m3/h Rotary vane pump with oil mist filter and gas ballast
Options & Accessories
- Accessories Options:
Moisture Sensor
Oxygen Sensor
Solvent Absorber
Heated Antechamber
Refrigerator
Cold Well
Microscope Unit
Ionizer without fan
Various Feedthroughs - Other Options:
Customized box sizes
Integrating customer’s instruments into the glovebox, like evaporator, ALD System and Spin Coater etc.
Polycarbonate Window(s)
Quick Release Window(s)
Antechamber with interlocking door;
Small chamber with sliding tray
Antechamber purging only without vacuum
Antechamber refilling from Gas cylinder instead of box
Rectangular Antechamber
Flexible shelves – adjustable up & down – adjustable back & forth
12m3/h Rotary pump
Dry scroll pump
Prompt service